188 OBD
TRIGGER CONTROL
5. ↓↑ Toggles the trigger slope (rising or falling). The current trigger
slope is indicated on the top right of the display.
↑= trigger on rising edge
↓= trigger on falling edge
6. TRIG There are three Trigger Modes: NORMAL, AUTO, and SINGLE
SHOT. Pressing 6 cycles through the three modes.
In NORMAL mode (the default mode), the Tester waits for the
trigger to occur before the waveform is displayed. Normal mode is
indicated by a "n" in the upper right corner of the display.
In AUTO mode, if a trigger does not occur for 250 ms, a trigger is
forced to occur. This allows signals to be found more easily since
the display shows the waveform even when a trigger does not
occur. Auto mode is indicated by an "a" in the upper right corner of
the display.
In SINGLE SHOT mode, the trigger is only activated when the
signal crosses the trigger level, or the E key is pressed.
While waiting for the trigger the single shot indicator shows an
upper case "S". When a trigger occurs, the indicator changes to a
lower case "s". The display remains frozen until the E key
is pressed, or a new trigger mode is chosen.
7. HOLD The HOLD mode freezes the current display so that the waveform
can be analyzed. The display is frozen until the 7 key is pressed
again, the trigger mode is changed, or the display is adjusted (time
scale or scale changed). When the hold mode is active, the right
Red LED is turned on.
Comentarios a estos manuales