SEKURE SX-3100LCD Manual del operador Pagina 362

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 378
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 361
350 GAS ANALYZER
INSTALLING NEW O2 SENSOR
The software indicates when the O2 (oxygen) sensor requires replacement. The normal
life expectancy is one year. The sensor degrades in the presence of oxygen. Because of
this, replacement sensors should remain in a sealed package until needed in order to
maximize sensor life.
To replace the O2S Sensor, do the following:
1. Precondition the sensor: Open oxygen sensor package 5 to 10 hours prior to
installation. If the sensor is calibrated immediately after exposure to air, it may drift.
2. While the Gas Analyzer is shut off, replace the old sensor with the new sensor;
remember to reconnect the O2S cable to the oxygen sensor.
3. Discard the old oxygen sensor.
4. Perform a Leak Test (5 from the <MAINTENANCE> menu) to check system
integrity.
5. Run the <INSTALL O2 SENS> by selecting 6 from the <MAINTENANCE> menu.
O
2
SENSOR
"O" RING
FIGURE 12. Replacing the Oxygen Sensor
Vista de pagina 361
1 2 ... 357 358 359 360 361 362 363 364 365 366 367 ... 377 378

Comentarios a estos manuales

Sin comentarios